The name F-center comes from the German word for color, “farbe,” as crystals containing these point defects are highly colored. The F-centers are point defects and can be readily formed in alkaline halides with the aid of a source of ionizing radiation, such as an X-ray source. Apparently, the ionizing radiation causes the loss of electrons from halide ions. An electron then becomes trapped in the halide ion vacancy. The color is the result of the absorption of a photon by the trapped electron and excitation from the ground state to an excited state for the F-center. This is a classic case of particle in a box.
Safety:
-Wear eye protection
Procedure:
Step 1. Prepare a microscope slide with an aluminum holder. Take a piece of tape and wrap it around the slide so that the sticky side is facing out. Place the aluminum holder on top of the slide with the tape centered in the hole. The end of the aluminum must match the end of the slide.
Step 2. Since the instrument is a powder x-ray diffractometer your sample should be a powder. Grind the sample using a rotary motion with a mortar and pestle. The finer the powder the better the data.
Step 3. Pour the powder on to the holder. Use another microscope slide to wipe off the excess. Best results are obtained if the sample is level and as deep as the holder.
Step 4. When you turn the latch to open the door the yellow light labeled “shutter open” will turn off. This is because turning the latch closes the shutter between the x-ray tube and the sample. With the shutter closed x-rays will not be present.
Make sure the plastic tray is under the sample This container is meant to catch any powder which could otherwise fall to the x-ray source and damage the machine. Hold both the slide and the aluminum holder. Insert the pair together into the clip holder.
Step 5. Run a powder diffraction scan of an alkali halide. (The sample must be tilted to get a direct beam on the sample and create F-centers.) At the end of the scan, open the door to release the shutter and block the x-rays. Immediately observe the color of the irradiated area.
Different alkali halides have different color F-centers. (F is an abbreviation for Farbe, the German word for color.)
– How long does it take, as illustrated by the loss of color, for the electron in the F-center to return to its origin? Do different materials take the same time?
Summary Questions:
- Predict the relative sizes of the KBr, KCl, and NaCl unit cells.
- The tallest peak is the 200 reflection. What is your measured size of the unit cell?
- What are the F-center colors for each of these?
- Does the spectral order of the colors match the unit cell sizes?
Materials:
- KCl
- NaCl
- KBr
- Powder X-ray Diffractometer (Rigaku Miniflex-II shown here.)